Find the time delays τL, τC, and τR so that, on transmit, the phased array will focus on the point F.

A phased array ultrasound transducer consists of three elements (L,C,R) each of dimensions h = 3 mm and d = 8 mm, as shown in Figure P11.3. Each element has a square face, s(x, y) = rect). In the first scenario [Figure P11.3(a)], the transducer has a silicone sheet of thickness s = 2 mm on its front face, which is placed on the skin. The acoustic impedances Z and the speed of sound c in silicone and in skin are given in the table below. For parts (a) to (e), assume that only one element, C, is generating and receiving ultrasound waves. The envelope of the wave is ne(t) = rectand its frequency is 2 Mhz.

(a) Sketch the A-mode signal from the time of the transmit pulse through the time of the echo from the silicone-skin interface. Assume a geometric field approximation and ignore attenuation and speckle. Clearly label key times and amplitudes.

In order to reduce the size of the echo from the skin, an acoustic gel of width g = 3 mm is introduced between the silicone and the skin, as shown in the figure [Figure P11.3(b)]. The gel is selected such that 

(b) Find the speed of ultrasound in the gel, cgel, so that there can be no refraction at the gel-skin interface regardless of the angle of incidence.

(c) Sketch the A-mode signal from the time of the transmit pulse through the time of the echo from the gel-skin interface. Assume a geometric field approximation and ignore attenuation and speckle. Clearly label key times and amplitudes.

(d) Suppose you want your imaging system to be insensitive to the echo from the skin. What should the sensitivity L (in dB) be?

In parts (e) and (f) ignore the silicone and gel.

(e) Is point F in Figure P11.4 in the near field or far field of the central transducer?

(f) Find the time delays τL, τC, and τR so that, on transmit, the phased array will focus on the point F.

find the cost of your paper

Explain why attenuation is not a big problem in PET.

Consider a 2-D object consisting of two triangle compartments, as shown in Figure P9.4. Suppose a solution containing a 511 KeV gamma ray emitting radionuclide with concentration f = 0.5….

Give the mean and the variance of the reconstructed image, mean[ˆ f(x, y)] and var[ˆ f(x, y)].

Ignoring the inverse square law and attenuation, an approximate reconstruction for SPECT imaging is given by where c˜() =  {||W()} and W() is a rectangular windowing filter that cuts off at = 0…..

Find the numerical responses in each to an event in crystal C(4, 6).

Suppose a PET detector comprises four square PMTs (arranged as a 2 by 2 matrix) and a single BGO crystal with slits made in such a way that it is….